<4. Setting converter>
4-18
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
4.4.4 Fail
A fail contact is energized when a fail is detected. When a fail is detected, the contact output is
l Process parameter slot
Select sensors from “Sensor 1-1”, “Sensor 1-2”, “Sensor 1-3”, “Sensor 1-4”, “Sensor 2-1”,
“Sensor 2-2”, “Sensor 2-3”, “Sensor 2-4.”
Fail is output only by the sensors you select. Fail from FLXA402 is always valid.
NOTE
When an internal communication error occurs, a fail is output. However, the error location would
l Setting
Select “Hard Fail only” or “Hard + Soft only”.
Hard Fail only: Contact output reacts only to Failure.
or “Maintenance Required”.
When Failure happens, contact output stays On (S1/S2/S3: close, S4: open).
4.4.5 USP
USP/EU limit (for only SC)
in USP <645> Water Conductivity Stage 1- Temperature and Conductivity Requirements, the
contact output turns On (S1/S2/S3: close).
l Process parameter slot
Determine sensors to detect USP. from “Sensor 1-1”, “Sensor 1-2”, “Sensor 1-3”, “Sensor 1-4”,
“Sensor 2-1”, “Sensor 2-2”, “Sensor 2-3”, “Sensor 2-4”.
For information on USP, see Appendix of Operation of SC, IM 12A01F01-03EN