<4. Setting converter>
4-12
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
4.4 Contact output settings
The following function can be assigned to the contact output.
The contact output has expected lifetime of around 1 million counts. This may vary depending on
contact mA or load capacity.
S4: Alarm, Error, Simulate, Fail safe
Contact AC/DC Power On,
Normal operation
AC/DC Power On,
Abnormal operation
AC/DC power orr
S1, S2, S3
NC
C
NO
NC
C
NO
NC
C
NO
S4
NC
C
NO
NC
C
NO
NC
C
NO
4.4.1 Alarm
point, direction, Hysteresis, Delay time, Expiry time.
l Process parameter slot
Read 4.3
l Process parameter
Set process parameter for Alarm. Select parameters for alarm from the following list.
For all measurement system: temperature
PH : pH, ORP, rH
SC : Conductivity TC1, Conductivity TC2, Resistivity 1,
Resistivity 2, Concentration 1, Concentration 2
DO Oxygen
the converter Script *1
DO, Average pH, Average ORP, Average CONDUCT, Average RESIST, Average DO,
*1 FLXA402 does not support script on Latest revision
*2 When multiple sensors are connected, the arithmetic starts on Sensor 1-1 and Sensor
2-1.
fails.