<4. Setting converter>
4-28
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
l PV, SV, TV, QV


NOTE
Latest revision does not support Script, or multiple sensor arithmetic.
Device variable are selected by sensor connection number. MODBUS address has to be
assigned on each connected sensor appropriately in advance. Read4.6.1 MODBUS
setting4.6.1 MODBUS setting.
NOTE
Latest revision allows the second sensor module to connect only one sensor. Do not select
Address CH 2-2, 2-3 or 2-4 to set Device variables. Should you select, AO current 1 would be
activated, but not would the expected operation.
NOTE

open a setting screen on the converter or gain an authorization of MODBUS setting, you cannot

Refer to TI 12A01F01-61 on HART communication.
4.6.3 Ethernet setting (E)
Modbus TCP/IP is available by Ethernet cable-connection between the converter and master
devices.
Ethernet
Maximum session times 2
Protocol Modbus//TCP
Port number 502
Ethernet 10/100 Mbps speed rate, Internet Protocol version 4 (Ipv4)
Set “DHCP enable” to “On” if you want DHCP server to assign IP address automatically. Set it
󰀨󰀨
When you use a static IP address, set “Subnet mask” and “Default gateway”, according to
Ethernet you use.
The next table shows default of Ethernet parameter.
Parameter name Default
DHCP 󰀨
IP address 192.168.1.10
Subnet mask 255.255.255.0
Default gateway 192.168.1.1