<4. Setting converter>
4-27
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
Example of MODBUS addressing
If you want to assign SENCOM SA - MODBUS address 8 to mA2,

mA2, go to “Converter menu” > mA output > mA2 > Process parameter slot, and select.
Figure 4.13 Address CH assignment and mA output
n MODBUS - Converter
FLXA402 as slave communicates with HOST devices, via MODBUS RS-485, Bluetooth, or
Ethernet (Modbus TCP/IP). Select a network system according to the hardware architecture you
use.
For further technical information on MODBUS, read TI 12A01F01-62EN.
First of all, before Bluetooth paring with devices, you need to create Bluetooth device name with
up to 12 characters. The default is serial number of the device.
On each device of the network, “FLXA402- [Bluetooth device name]” is diaplayed.
[Bluetooth device name] is loaded on to the system display after the FLXA402 reboots.
4.6.2 HART setting

NOTE
 

l Loop Current Mode
For Point-to-Point, stay on the default “Enable”.

“Enable” mA1 output normal Loop current
 
l Network address
Point-to-Point stay on the default “0”.
 󰀨
