<4. Setting converter>
4-13
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
l Setpoint, Direction, Hysteresis, Delay time, Expiry, time
Setpoint; This decides the value to activate contact output (S1/S2/S3: close, S4: open)
when an alarm occurs.
Direction: This selects “High” or “Low” to report the value is outside of the limit.
 
 
beyond the Setpoint in Direction.
Setpoint
Hys.
SC
off on off
Delay time Delay time
t (sec)
Figure 4.5 Alarm contact
l Expiry time
󰀨
fails. To cancel the alarm, go to the Error code and cancel the error. If you set Expiry time to 0.0
second, it turns invalid.
Figure 4.6 Error code example when a contact output exceeds the Expiry time