<2. Installation >
2-3
IM 11M10A01-01E
Mode Sequence
MODE
ENT.
20.6
ppm
%
s
d
SET
AIR
1
000ppm
4-20
mA
Measurement mode
(Normal measurement)
Oxygen concentration is
displayed on main display.
Setting mode
(Alarm setting)
Can set alarm
in range 0 to 25% O
2.
Switch to setting mode
NA
NA
Press both to
switch to 4-20 mA
output range
NA
Cal1 mode
Atmosphere
(20.6% O
2)
Calibrate in atmosphere.
Cal2 mode
(1000 ppm)
Calibrate with1000 ppm O
2
for better accuracy at low
concentrations.
Switch to Cal1 mode
Increase setting
Decrease setting
Enters setting and
switches to measure.
Switch to Cal2 mode
NA
NA
Switch to measure
Starts calibration.
At end, switches to
measure.
Increase setting
Decrease setting
Starts calibration.
At end, switches to
measure.
Key SW
StateTransitionFlowChart(Example)
MODE
ENT.
SET
AIR
1
000ppm
4-20
mA
ENT.
ENT.
ENT.
MODE
MODE
MODE
Setting mode
Calibration 2 mode
20.6
ppm
%
s
d
(Example)
NA
NA (Calibrate)
(Calibrate)
Increase setting Decrease setting
Increase setting Decrease setting
(Enter setting)
(Press simultaneously)
4-20 mA output range switching
Calibration 1 mode
Measurement mode
T2.2E.EPS