14-21
IM 760401-01E
Communication Commands 2 (System of Commands Complying to the IEEE 488.2-1992 Standard)
14
[CONFigure]:MODE
Function Sets the measurement mode of current and
voltage, queries the current setting.
Syntax [CONFigure]:MODE {RMS|VMEan|DC}
[CONFigure]:MODE?
Example [CONFIGURE]:MODE RMS
[CONFIGURE]:MODE? -> :CONFIGURE:
MODE RMS
[CONFigure]:SCALing?
Function Queries all settings relating to the scaling
function.
Syntax [CONFigure]:SCALing?
Example [CONFIGURE]:SCALING? -> :CONFIGURE:
SCALING:STATE 0;PT:
ELEMENT1 1.000E+00;
ELEMENT2 1.000E+00;
ELEMENT3 1.000E+00;
:CONFIGURE:SCALING:CT:
ELEMENT1 1.000E+00;
ELEMENT2 1.000E+00;
ELEMENT3 1.000E+00;:CONFIGURE:
SCALING:SFACTOR:ELEMENT1 1.000E+00;
ELEMENT2 1.000E+00;
ELEMENT3 1.000E+00
[CONFigure]:SCALing:{PT|CT|SFACtor}?
Function Queries all scaling constants related to
{voltage|current|power}.
Syntax [CONFigure]:SCALing:
{PT|CT|SFACtor}?
Example [CONFIGURE]:SCALING:PT?->
:CONFIGURE:SCALING:PT:
ELEMENT1 1.000E+00;
ELEMENT2 1.000E+00;
ELEMENT3 1.000E+00
[CONFigure]:SCALing:{PT|CT|SFACtor}[:ALL]
Function Sets the scaling constants for all elements of
{voltage|current|power} at once.
Syntax [CONFigure]:SCALing:{PT|CT|
SFACtor}[:ALL] {<NRf>}
{<NRf>}=0.001 to 9999
Example [CONFIGURE]:SCALING:PT:ALL 1.000
Description The setting values are rounded as follows.
Less than 1.000 : Three digits after the decimal
point are valid.
1.000 to 9999 : The first five digits are valid.
[CONFigure]:SCALing:{PT|CT|SFACtor}:
ELEMent<x>
Function Sets the scaling constant for
{voltage|current|power} of each
element, queries the current setting.
Syntax [CONFigure]:SCALing:{PT|CT|
SFACtor}:ELEMent<x> {<NRf>}
[CONFigure]:SCALing:{PT|CT|SFACtor}:
ELEMent<x>?
<x>= 1 (WT210 single-phase
model)
1, 3 (WT230 three-phase,
three-wire model)
1 to 3 (WT230 three-phase,
four-wire model)
{<NRf>}=0.001 to 9999
Example [CONFIGURE]:SCALING:PT:
ELEMENT1 1.000
[CONFIGURE]:SCALING:PT:ELEMENT1?
-> :CONFIGURE:SCALING:PT:
ELEMENT1 1.000E+00
Description The setting values are rounded as described at
[CONFigure]:SCALing:{PT|CT|
SFACtor}[:ALL]
[CONFigure]:SCALing[:STATe]
Function Sets scaling ON/OFF, queries the current
setting.
Syntax [CONFigure]:SCALing[:
STATe] {<Boolean>}
[CONFigure]:SCALing:STATe?
Example [CONFIGURE]:SCALING:STATE OFF
[CONFIGURE]:SCALING:STATE?
-> :CONFIGURE:SCALING:STATE 0
[CONFigure]:SYNChronize?
Function Sets the measurement synchronization source/
queries the current setting.
Syntax [CONFigure]:SYNChronize {VOLTage|
CURRent|OFF}
[CONFigure]:SYNChronize?
Example [CONFIGURE]:SYNCHRONIZE VOLTAGE?
[CONFIGURE]:SYNCHRONIZE? ->
:CONFIGURE:SYNCHRONIZE VOLTAGE
[CONFigure]:VOLTage?
Function Queries all settings relating to voltage range.
Syntax [CONFigure]:VOLTage?
Example [CONFIGURE]:VOLTAGE? -> :CONFIGURE:
VOLTAGE:RANGE 600.0E+00;AUTO 1
[CONFigure]:VOLTage:AUTO
Function Sets the voltage auto range ON/OFF, queries
the current setting.
Syntax [CONFigure]:VOLTage:AUTO {<Boolean>}
[CONFigure]:VOLTage:AUTO?
Example [CONFigure]:VOLTage:AUTO ON
[CONFIGURE]:VOLTAGE:AUTO?
-> :CONFIGURE:VOLTAGE:AUTO 1
[CONFigure]:VOLTage:RANGe
Function Sets the voltage range/queries the current
setting.
Syntax [CONFigure]:VOLTage:RANGe {<voltage>}
[CONFigure]:VOLTage:RANGe?
When the crest factor is set to 3
<voltage>=15V to 600V (15, 30, 60,
150, 300, 600V)
When the crest factor is set to 6
<Voltage> = 7.5 V to 300 V (7.5, 15,
30, 75, 150, or 300)
14.3 Commands