14-32 IM 760401-01E
Data Format/Output Format of Normal and Harmonic Measurement Data <ASCII>
The data format/output format of normal and harmonic measurement data which is
requested by
MEASure[:NORMal]:VALue? or MEASure:HARMonics:VALue?, is as follows.
Data Format of Normal Measurement Data
All data of the <harmonic measurement function> are output in the <NR3> format.
(Example) 123.45E+00
V,A,W,VA,VAR,PF,DEGR,VHZ,AHZ,VPK,APK,MATH mantissa: max. 5 digits +
exponent: 2 digits
WH,WHP,WHM,AH,AHP,AHM mantissa: max. 6 digits + exponent: 2 digits
(max. 5 digits in case of negative value)
The sign of the mantissa will only be applied in case of negative values. However,
phase lead and lag (in case of phase angle (DEG)) will be shown as follows.
LEAD
+180.0E+00
LAG -180.0E+00
in phase 0.0E+00 (The mantissa will be proceeded by a space)
In case of overrange or computation over,“9.9E+37“(+) will be output.
(i.e. in case the display shows –oL–, –oF–, PFErr, dEGEr, ErrLo, or ErrHi)
In case no data is present (i.e. the display shows – – – – –),
“9.91E+37“ (NAN) will be
output.
The elasped integration time is output as hours, minutes, seconds in the <NR1>
format.
(Example) 999,59,59
Only one of the six parameters, VHZ1, VHZ2, VHZ3, AHZ1, AHZ2, or AHZ3, can be
measured for frequency. Data that is not measured is set to no data (9.91E+37).
VHZΣ, AHZΣ, VPKΣ, and APKΣ always output no data (9.91E+37).
Output Format of Normal Measurement Data
The communication output is set ON by any of the commands starting with
MEASure[:NORMal]:ITEM” and the normal measured/computed data or elapsed
integration time are output according to the following order of priority. Besides, in
case of recalling normal measurement or integration data, the data number will be
output in <NR1> format as well. Data will be output in the following order
corresponding to each element. However, note that for model 760401 only element 1
is valid, and for model 760502 only element 1, 3 and S are valid.
(
0. Data number in case of recalling)
1. V1 V2 V3 V
2. A1 A2 A3 A
3. W1 W2 W3 W
4. VA1 VA2 VA3 VA
5. VAR1 VAR2 VAR3 VAR
6. PF1 PF2 PF3 PF
7. DEGR1 DEGR2 DEGR3 DEGR
8. VHZ1 VHZ2 VHZ3 VHZ
9. AHZ1 AHZ2 AHZ3 AHZ
10. WH1 WH2 WH3 WH
11. WHP1 WHP2 WHP3 WHP
12. WHM1 WHM2 WHM3 WHM
13. AH1 AH2 AH3 AH
14. AHP1 vAHP2 AHP3 AHP
14. AHM1 AHM2 AHM3 AHM
16. TIME (elapsed integration time)
Each data is divided by a comma”
,” and is ended by the terminator <RMT>.
14.3 Commands

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA WT210 and is the answer not in the manual?

YOKOGAWA WT210 Specifications

General IconGeneral
BrandYOKOGAWA
ModelWT210
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals