EasyManua.ls Logo

YOKOGAWA YTA Series - 7. General Specifications (Continued)

YOKOGAWA YTA Series
61 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-4
IM 01C50C05-01EN
Input entity parameters: Ui(Vmax)=30V, Ii(I max)=120mA, Pi=0.84W,
Ci=1nF, Li=10
μ
H
Output entity parameters: Uo(Uoc)=9.6V, Io(Isc)=28mA,
Po(Pmax)=67.2mW, Co(Ca)=3.5
μ
F, Lo(La)=35mH
7.3 EMC Conformity
EN 61326-1 Class A, Table2
EN 61326-2-3
CAUTION
This instrument is a Class A product, and it is designed for use in the
industrial environment. Please use this instrument in the industrial
environment only.
7.4 EURoHSDirective
Applicable standard: EN IEC 63000
Applicable production sites are shown below
The condition of the RoHS compliant production site are as follows:
Singapore, China
The production sites can be conrmed by the serial number shown in the
frame of “No.” in the name plate of the product.
Serial numbers (9 letters): AAnnnnnnn
AA: Identication code of production site
Singapore: Use “C2” or “C0” China: Use “S5”

Related product manuals