IM 01C50B01-01E
7-4
7. STANDARD SPECIFICATIONS
Table 7.2 YTA110 Effect of Ambient Temperature
Sensor Type
T/C B
T07021.EPS
Temperature Coefficient
Thermocouples E, J, K, N, T, L, U
Thermocouples R, S, W3, W5
100C Reading < 300C
300C Reading
RTD
mV
ohm
0.08C + 0.02% of abs.reading
0.25C + 0.02% of abs.reading
1C + 0.02% of abs.reading
0.5C + 0.02% of abs.reading
0.08C + 0.02% of abs.reading
0.002 mV + 0.02% of abs.reading
0.1 + 0.02% of abs.reading
Table 7.3 YTA310, YTA320 Effect of Ambient Temperature
Sensor Type
Input Range
A/D Coefficient D/A Coeffieicient
C
F
T/C
B
212 to
572
± (
0.530 C0.080 % of reading)
572 to
1832
± (
0.350 C0.021 % of reading )
1832 to
3308
± (
0.140
C)
E
-328 to
1832
± (
0.035 C0.042 % of abs.reading)
J
-328 to
32
± (
0.039 C0.020 % of abs.reading)
32 to
2192
± (
0.039
C0.0029 % of reading)
K
-328 to
32
± (
0.046 C0.020 % of abs.reading)
32 to
2502
± (
0.046
C0.0054 % of reading)
N
-328 to
32
± (
0.054 C0.010 % of abs.reading)
± {0.0088% of span0.007% of (readingLRV)}
32 to
2372
± (
0.054
C0.0036 % of reading)
R
-58 to
392
± (
0.210 C0.032 % of abs.reading)
392 to
3214
± (
0.150
C)
S
-58 to
392
± (
0.210 C0.032 % of abs.reading)
392 to
3214
± (
0.150
C)
T
-328 to
32
± (
0.046 C0.036 % of abs.reading)
32 to
752
± (
0.046
C)
W3
32 to
2552
± (
0.100 C0.0040 % of reading)
2552 to
4172
± (
-0.130
C0.020 % of reading)
W5
32 to
2552
± (
0.100 C0.0040 % of reading)
2552 to
4172
± (
-0.120
C0.020 % of reading)
L
-328 to
32
± (
0.039 C0.020 % of abs.reading)
32 to
1652
± (
0.039
C0.0029 % of reading)
U
-328 to
32
± (
0.046 C0.036 % of abs.reading)
32 to
1112
± (
0.046
C)
RTD
Pt100
-328 to
1562
± (
0.047 C0.009 % of reading)
Pt200
-328 to
1562
± (
0.065 C0.012 % of reading)
Cu
± ( 0.320 C0.120 % of reading)-94 to 302
mV
± (0.001mV0.0043 % of abs.reading)
ohm
± (
0.040 0.0088 % of reading)
T07022.EPS
100 to
300
300 to
1000
1000 to
1820
-200 to
1000
-200 to
0
0to
1200
-200 to
0
0to
1372
-200 to
0
0to
1300
-50 to
200
200 to
1768
-50 to
200
200 to
1768
-200 to
0
0to
400
0to
1400
1400 to
2300
0to
1400
1400 to
2300
-200 to
0
0to
900
-200 to
0
0to
600
-200 to
850
-200 to
850
-70 to 150
Pt500
JPt100
-328 to
1562
± (
0.047 C0.009 % of reading)
-328 to
932
± (
0.047 C0.009 % of reading)
-200 to
850
-200 to
500
Ni120
-94 to
608
± (
0.016 C0.007 % of reading)
-70 to
320
Note: Temperature Effect = A/D coeffieicnt + D/A coefficient (The data in the table is the coeffcient per 10C change.)
Example 1; Pt100, 0 to 200C calibration range, 50C reading
(0.047C 50C 0.009%) [200C 0.0088% (50 0) 0.007%]
= (0.047C 0.0045C) (0.0176C 0.0035C )
= ± 0.0726C [ per 10C change ]
Example 2; T T/C, 100 to 100C calibration range, 250C reading
(0.046C | 250C | 0.036%) {200C 0.0088% [250 (100)] 0.007%}
= (0.046C 0.018C) (0.0176C 0.0035C )
= ± 0.0851C [ per 10C change ]