<8. Detailed Data Setting>
8-2
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
5
15
25
35
45
55
65
75
85
95
5 15 25 35 45 55 65 75
Minimum oxygen concentration, vol%O
2
(for a minimum current of 4 mA)
Minimum-Maximum setting range of oxygen concentration
Maximum oxygen concentration, vol%O
2
(for a maximum current of 20 mA)
Ranges over which oxygen
concentrations can be set
Outside ranges
F8.0E.ai
Figure 8.1
l Humidity (amount-of-moisture-content) and relative humidity setting range
The minimum humidity is set to 0% H
2
O or ranges from 26 to 100% H
2
O. The maximum humidity
ranges from 25% to 100% H
2
O, and must be greater than 0.8 times plus 23 the humidity set for
the minimum. See an example shown below or the graph Figure 8.2.

H
2
O with % RH
Setting example 1
If the minimum setting (for a 4 mA point) is 0% H
2
O, you must set the maximum (20 mA) point at
more than 25% H
2
O.
Setting example 2
If the minimum setting (for a 4 mA point) is 26% H
2
O, you must set the maximum (20 mA) point
at more than 44% H
2
O. (26 x 0.8 + 23 = 44 vol% H
2
O). (Numbers after the decimal point are
rounded up.

Table of Contents

Other manuals for YOKOGAWA ZR802G

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ZR802G and is the answer not in the manual?

YOKOGAWA ZR802G Specifications

General IconGeneral
BrandYOKOGAWA
ModelZR802G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals