<10. Other Functions>
10-22
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
Table 10.9 Initialization Items and Default Values (Oxygen Analyzer)
Item Initialization Parameter Default setting
Model Setting - Not initialized
Language - Not initialized
Display settings Display item 1st display item selection Oxygen
concentration
2nd display item selection Output item 1
3rd display item selection Favorite
Tag name Deleted
Trend graph Graph selection items Oxygen
concentration
Graph sample time 15min.
Set graph limits High limit 25.0%O2
Low limit 0.0%O2
Others setting Home screen auto return time Disable
NE107 mode 󰀨
Backlight time Disable
Logbook settings All ON
Favorite settings Favorite display item 1 Sensor details
screen
Favorite display item 2 Converter details
screen
Favorite display item 3 Trend screen
Favorite display item 4 No setting
Communication
settings
MODBUS setting MODBUS address Not initialized
Baud rate Not initialized
Parity Not initialized
TCP settings IP Address Not initialized
Subnet mask Not initialized
Default gateway Not initialized
DHCP enable Not initialized
Calibration settings Calibration mode Manual
Automatic calibration procedure Span and Zero
Zero gas concentration 1.00%O2
Span gas concentration 21.00%O2
Calibration time
setting
Hold time [mm:ss] 10:00
Calibration time [mm:ss] 10:00
Interval [day] 30
Interval [hour] 0
First start date [YY/MM/DD] 00/01/01
First start time [hh:mm] 00:00
Blowback setting Mode Mode None
Blowback time Hold time [mm:ss] 10:00
Blowback time [min] 10
Blowback time [sec] 0s
Interval [day] 30
Interval [hour] 0
First start date [YY/MM/DD] 00/01/01
First start time [hh:mm] 00:00