Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4Setup
24 IM/AX4PH Rev. P
Calibrate Buffer 2 (Two-Point Calibration only)
Immerse Sensor A in the second buffer solution.
Press the key to initiate calibration.
Note. To abort calibration, press the key again at any time before calibration is
complete – see previous page.
The center display line shows the measured sensor output in millivolts.
As calibration proceeds, a progress indicator appears in the lower display line. When
the measured sensor output stabilizes, the lower display line shows
##### 100 % #####.
The display then changes for 2 seconds to show the temperature-corrected buffer
value in the upper display line, then advances automatically to the next frame.
Calibration Message
See Table 4.1 for details of calibration messages.
Slope Value
% slope value.
A value between the programmed minimum % slope value (see Set Min Slope in the
CONFIG. SENSORS page – Section 5.3) and 105 % is displayed. If the value is outside
these limits, check the electrode system.
pH check value.
Displayed as an additional indication of electrode system condition; 7 ph is the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
Sensor B (dual input analyzers only) calibration is identical to Sensor A calibration.
Probe Type for Sensor B (dual input analyzers only) set to Redox or ORP and Enable
Cals.
set to No (Section 5.3) – return to top of page.
Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28.
Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29.
SECURITY CODE
CONFIG. DISPLAY
Sensor Cal. A
A: Immerse Buf.2
9.18
pH
20.0
Deg.C
#### 100% ####
128
mV
A: Calibration
-----
Passed
A: Slope & Check
100.0
%
7.00
pH
A: Cal. Buffer 2
4.01
pH
A: Buffer Method
set to Auto 1-Pt
A: Buffer Method
set to Auto 2-Pt
Sensor Cal. B
SENSOR CAL.