Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6Installation
44 IM/AX4PH Rev. P
6 Installation
6.1 Siting Requirements
Note.
Mount in a location free from excessive vibration, and
where the temperature and humidity specification will
not be exceeded.
Mount away from harmful vapors and/or dripping
fluids and ensure that it is suitably protected from
direct sunlight, rain, snow and hail.
Where possible, mount the analyzer at eye level to
allow an unrestricted view of the front panel displays
and controls.
Fig. 6.1 Siting Requirements
C – Within Environmental Limits
B – Within Temperature Limits
A – Maximum Distance Between Analyzer and Sensor
Sensor
Maximum Distance
100m (275 ft)
55°C
(131°F)
Max.
–20°C
(–4°F)
Min.
IP65*
* Refer to Specification, page 64