EasyManuals Logo

ABB EasyLine EL3000 User Manual

ABB EasyLine EL3000
216 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #111 background imageLoading...
Page #111 background image
EL3000 CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D 111
Notes on the operating concept
The operating concept of the gas analyzer requires that the functions needed in normal operating mode are operated and configured
directly on the device.
On the other hand, those functions that are seldom needed, for example when commissioning the device, are named, configured and
then loaded into the gas analyzer offline using the ECT ‘EasyLine Configuration Tool’ software tool, also referred to as the
‘Configurator’ in this instruction.
Overview of functions
Function
Device Configurator Modbus
Automatic Calibration
Start/cancel automatic calibration (can also be done via digital inputs)
X X
Power
-up/power-down a cyclically timed automatic calibration,
see Automatic calibration: Control on page 118
X X X
Cycle time of automatic calibration.
X
End-point calibration together with zero point calibration
X
X
Date and time of next automatic calibration (start of cycle)
X X
Test gas concentration, see Set test gas concentration on page 117
X
X
X
Purge time
X
X
Output current behavior (for automatic and manual calibration)
X
Calibration method (Magnos206, Magnos28)
X
X
Pump on during calibration
X
Manual Calibration
Calibration method
X
X
Test gas concentration, see Set test gas concentration on page 117
X
X
Perform calibration
X
Maintenance functions
Pump control, see Operate pump on page 198
X
X
Calibration reset, see Perform calibration reset on page 178
X
Calibrate pressure sensor / set air pressure value, see Pressure correction on page 194
X
Measuring the calibration cell, see Uras26, Limas23 measuring the calibration cell on page 179
X
Drift, Delta drift, see Drift indicator on page 194 (display)
X
X
So
ftware version, see Instrument information on page 198 X X
Status information, see Possible status messages on page 168
X
X
Component parameters
Measuring range parameters, (see Measuring range switchover on page 141)
X
X
Limit value parameters, see Limit values on page 147
X
Low pass time constant, T90 time, filter, see
Set low-pass time constant on page 148 X X
Active component, (see Active component selection on page 149)
X
X
X
Measuring range, (see
Sample components on page 150) X
Modbus parameters*
X
PROFIBUS parameters**
X
Ethernet parameters, see Setting the IP address on page 112
X
X
Signal inputs and outputs, see Configuring signal inputs and outputs on page 152
X
* You will find detailed information regarding Modbu in the interface description ‘COM/EL3000/MODBUS’.
** You will find detailed information regarding PROFIBUS® in the technical information ‘30/24-415’.

Table of Contents

Other manuals for ABB EasyLine EL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB EasyLine EL3000 and is the answer not in the manual?

ABB EasyLine EL3000 Specifications

General IconGeneral
BrandABB
ModelEasyLine EL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals