EL3000 CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D 173
Troubleshooting
Note
For the Fidas24 analyzer, the instructions under Fidas24
Troubleshooting on page 174 must be observed.
Flow error
Gas lines or filter contaminated, clogged up or leaky
disconnect the gas analyzer from the gas treatment
system.
blow out the lines with compressed air or unblock them
mechanically.
replace filter inserts and fill material.
check gas lines for leaks.
Gas paths in the gas analyzer kinked or leaky
disconnect the gas analyzer from the gas treatment
system.
Check whether the gas lines in the gas analyzer are kinked
or have become detached from the connections.
Check the gas feed paths in the gas analyzer for leak
tightness (see Checking gas path leak tightness on
page 199).
Outlet pressure not the same as atmospheric pressure
Make sure that the outlet pressure is the same as the
atmospheric pressure.
Guide exhaust gases directly into the atmosphere or
through a line with a large internal diameter which is as
short as possible, or into a gas discharge line.
Do not guide exhaust gases via restrictions or shut-off
valves.
Unstable display of measured value
Vibrations
Provide measures to reduce the vibrations.
Increase the low pass time constant (see Set low-pass
time constant on page 148).
Leaks in the gas feed paths
Check the gas feed paths in the gas analyzer for leak
tightness (see Checking gas path leak tightness on
page 199).
Loss of sensitivity
Check amplification drift display of the sample
component (see Drift indicator on page 194).
Uras26: Emitter modulation uneven
Have the emitter and modulator checked by Service.
Limas23: emission intensity of UV lamp too low or sample
cell contaminated
Check the intensity display (see Limas23 Intensity
display on page 179).
Replace the UV lamp (see Limas23Replacing the UV
lamp (EDL) on page 183) or clean the sample cell (see
Limas23Clean the sample cell on page 180).