EL3000 CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D 161
Configuration of QAL3 monitoring
NOTICE
Data loss of the QAL3 data!
The QAL3 data stored on the memory card will be lost if the
gas analyzer configuration is changed using the ECT or
Optima TCT Light software tools, or if the memory card is
mounted in another gas analyzer.
In these cases the QAL3 data must be exported first (see
Exporting or deleting QAL3 data on page 163).
Note
By default set at the factory, QAL3 monitoring is disabled for
all measured components, i.e. no QAL3 data is stored.
If a password (see Password on page 156) is configured in
the gas analyzer, this password must also be entered when
configuring the QAL3 monitoring.
Opening the QAL3 monitoring web interface
1. Make sure that the gas analyzer is connected to a PC via the
Ethernet interface.
2. Open the web browser on the PC, enter the IP address of the
gas analyzer (if necessary, contact the network
administrator) and establish the connection.
The web interface of the QAL3 monitoring of EasyLine
EL3000 is open.
Configuration of sample components
1. In the QAL3 monitoring web interface, click ‘QAL3
Monitoring’.
2. On the ‘QAL3 Monitoring’ page, click ‘Components’ (sample
components).
3. On the ‘Components’ page, click on the sample component
to configure.
4. On the ‘Edit Component’ page (edit sample components),
enter the following information and values:
Active’ (QAL3 monitoring of the sample component
enabled),
Installation’ (installation location),
Technician’ (technician),
Comment’ (comment),
sAMS Zero’ (s
AMS
-value for zero point),
sAMS Span’ (s
AMS
-value for reference point).
5. Leave the page by clicking ‘Save’ (save) or ‘Cancel’ (cancel).
Configuration of pressure output
1. In the QAL3 monitoring web interface, click ‘QAL3
Monitoring’.
2. On the ‘QAL3 Monitoring’ page, click on ‘Settings’ (settings).
3. In the ‘Printing’ window (print output), check and if necessary
change the number of lines on the first page (‘Number of
lines on first page recommendation for high format: 35)
and on the following pages, (‘Number of lines on all other
pagesrecommendation for high format: 50).
4. Leave the page by clicking ‘Save’ (save) or ‘Cancel’ (cancel).

Table of Contents

Other manuals for ABB EasyLine EL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB EasyLine EL3000 and is the answer not in the manual?

ABB EasyLine EL3000 Specifications

General IconGeneral
BrandABB
ModelEasyLine EL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals