15
IC2
IC+
IC –
IC+ IC– IC2 IC–
HART connections
where applicable
MagMaster Transmitter
Common
+ve
Normal
Alternative
A
1
1
0
0
.
0
°
C
B
O
I
L
E
R
6
T
E
M
P
E
R
A
T
U
R
E
HART link
IC2
IC+
IC –
A
1
1
0
0
.
0
°
C
B
O
I
L
E
R
6
T
E
M
P
E
R
A
T
U
R
E
IC– IC2
HART connections
(where applicable)
MagMaster Transmitter
Common
Reverse +ve + HART
A
1
1
0
0
.
0
°
C
B
O
I
L
E
R
6
T
E
M
P
E
R
A
T
U
R
E
Receiver No. 2 – Reverse
Receiver No. 1 – Forward
Forward +ve
Information. For Multidrop HART installations, remove 'HART Link' and connect HART
systems directly to IC2: this allows the analog output function to be retained.
3 ELECTRICAL INSTALLATION…
3.4.5 Current Output – Fig. 3.20 and 3.21
Information.
• Output is fully programmable – see Programming Guide.
• Output is electrically separated from all other MagMaster connections.
• External isolators are not normally required and may significantly limit accuracy if used.
Fig. 3.20 Current Output Connections: Standard
Fig. 3.21 Current Output Connections: Dual Current Option
Information. Multidrop
HART mode cannot be used
with this configuration.