EasyManua.ls Logo

Abbott CentriMag - Page 119

Abbott CentriMag
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CentriMag
TM
Circulatory Support System Operation Manual (US) Page 119 of 126
© 2019 AbbottDocument No PL-0047, Rev 11 (December 2019)
2
nd
GENERATION CENTRIMAG CONSOLE
PARAMETER SPECIFICATIONS
Max. Product Life 5 years
MAG MONITOR
PARAMETER SPECIFICATIONS
DC Power 12 VDC, 12 W (from Console)
Dimensions
Height: 26.2 cm / 10.3 in
Width: 33.8 cm / 13.3 in
Depth: 5.4 cm / 2.1 in
Weight
2.0 kg / 4.4 lbs
Screen Size 12.1’’
Screen Resolution 800(H) x 600(V) pixels
Language Options English, Dutch, French, German, Italian and Spanish
Electrical Safety
Earth leakage current: < 500 µA
Touch current: < 100 µA
Patient leakage current: < 10 µA
Shipping Condition
Temperature: -29
o
C to 60
o
C / -20
o
F to 140
o
F
Relative humidity: 0% to 85%
Atmospheric pressure: 210 hPa 1100 hPa (157 mmHg
825 mmHg)
Operational & Storage
Conditions
Temperature: 10
o
C to 30
o
C / 50
o
F to 86
o
F
Relative humidity: 30% to 75%
Atmospheric pressure: 702 hPa 1100 hPa (527 mmHg
825 mmHg)
Monitor Application
Software
Version MCM3.01
IPL:
SW-0058-01
Rev 01
Application:
SW-0060-01
Rev 01
CPLD:
SW-0061-01
Rev 01
Max. Product Life 5 years
USB Port
Use only USB-compatible Memory Sticks
(FAT/FAT32 formatted)

Table of Contents