CentriMag
TM
Circulatory Support System Operation Manual (US) Page 119 of 126
© 2019 Abbott– Document No PL-0047, Rev 11 (December 2019)
2
nd
GENERATION CENTRIMAG CONSOLE
PARAMETER SPECIFICATIONS
Max. Product Life 5 years
MAG MONITOR
PARAMETER SPECIFICATIONS
DC Power 12 VDC, 12 W (from Console)
Dimensions
Height: 26.2 cm / 10.3 in
Width: 33.8 cm / 13.3 in
Depth: 5.4 cm / 2.1 in
Weight
2.0 kg / 4.4 lbs
Screen Size 12.1’’
Screen Resolution 800(H) x 600(V) pixels
Language Options English, Dutch, French, German, Italian and Spanish
Electrical Safety
Earth leakage current: < 500 µA
Touch current: < 100 µA
Patient leakage current: < 10 µA
Shipping Condition
Temperature: -29
o
C to 60
o
C / -20
o
F to 140
o
F
Relative humidity: 0% to 85%
Atmospheric pressure: 210 hPa – 1100 hPa (157 mmHg –
Operational & Storage
Conditions
Temperature: 10
o
C to 30
o
C / 50
o
F to 86
o
F
Relative humidity: 30% to 75%
Atmospheric pressure: 702 hPa – 1100 hPa (527 mmHg –
Monitor Application
Software
Max. Product Life 5 years
USB Port
Use only USB-compatible Memory Sticks
(FAT/FAT32 formatted)