CentriMag
TM
Circulatory Support System Operation Manual (US) Page 118 of 126
© 2019 Abbott– Document No PL-0047, Rev 11 (December 2019)
12.2 Appendix II – Technical Specification
2
nd
GENERATION CENTRIMAG CONSOLE
PARAMETER SPECIFICATIONS
AC Power
100 – 240 VAC
at 50/60 Hz, 170 VA
Battery Type & Chemistry Rechargeable internal battery, Lithium Ion
Battery Voltage 14.8 Volts
Available Battery Run
Time
approx. 120 min @ 3,500 RPM, 5.5 LPM
Battery Recharge Time 4 hrs to 90% charge, 5 hrs to 100% charge
Dimensions
Height: 10.0 cm / 3.9 in
Width: 26.6 cm / 10.5 in
Depth: 33.0 cm / 13.0 in
Weight 5.8 kg / 12.8 lbs
Pump Speed Range 0 – 5,500 revolutions per minute (RPM)
Pump Flow Range 0.0 – 10.0 liters per minute (LPM)
Flow Range Display -2.0 – 10.0 liters per minute (LPM)
Flow Resolution 10 mLPM
Pressure Range Display -150 – 900 mmHg
Alarm/Alert Audio Volume 70 dB
Language Options English, Dutch, French, German, Italian and Spanish
Electrical Safety
Earth leakage current: < 500 µA
Touch current: < 100 µA
Patient leakage current: < 10 µA
Fuse “5 x 20 mm, T 3.15A H 250V”
Shipping Condition
Temperature: -29
o
C to 60
o
C / -20
o
F to 140
o
F
Relative humidity: 0% to 85%
Atmospheric pressure: 210 hPa – 1100 hPa (157 mmHg –
Operational & Storage
Conditions
Temperature: 10
o
C to 30
o
C / 50
o
F to 86
o
F
Relative humidity: 30% to 75%
Atmospheric pressure: 702 hPa – 1100 hPa (527 mmHg –
2
nd
Generation CentriMag
Console Application
Software