7.18 Setting the Console Speed Step Resolution
Pressure Display
Speed Step Resolution
Flow Range
Flow Limit Sensitivity
Flow Recorder Speed*
Set RPM
Exit
Menu
ActiveInactive
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Menu
Menu
Menu
Menu
Menu
Menu
Normal
(100 rpm)
Fine
(50 rpm)
Set RPMMenu R
NormalSensitive
Set RPM
Menu Next
Normal
(0 – 10 lpm)
Reduced
(0 – 3 lpm)
Slow
(2 min)
Fast
(20 sec)
Set RPMMenu
L
Select
Enter
Return
* Monitor only
Extended Menu
Set RPMMenu
[ |Confirm]
[Execute|Cancel]
Set RPMMenu
[ |Confirm][Start|Cancel]
Copy Data*
Management*
Support Type
Language
Data Logger*
Figure 46: Monitor MENU Structure
Speed Step Resolution
The System may be used to treat patients with a wide range of body sizes and conditions.
The SPEED STEP RESOLUTION function is provided to allow the user to select smaller
incremental changes in speed (50 RPM increments) for small individuals, and larger (100
RPM) changes for the larger patients. The default setting is the 100 RPM option. To
access the SPEED STEP RESOLUTION function, depress the MENU keypad and scroll
down to the EXTENDED MENU option, then depress the Menu key until the SPEED
STEP RESOLUTION option appears. Choose between the STEP 50 (50 RPM) and STEP
100 (100 RPM) options using the UP and DOWN arrows. Once selected, this will dictate
the speed that the Pump will increase with each depression of the UP or DOWN ARROW
keypads when using the SET RPM option.