6 SPECIFICATIONS AND GENERAL DESCRIPTION
This section includes the product specifications and physical attributes of the System.
6.1 Classification
Table 4: System Classification
SYMBOL CLASSIFICATION DESCRIPTION
Type CF Defibrillator Proof
Equipment Type for protection
against electric shock.
None
Class 1 and internally powered
Equipment Classification for
protection against electric shock.
None
Continuous
Mode of Operation.
None Not for AP or APG
Not suitable for use in the
presence of a flammable
anesthetic mixture.
None
EtO for Pump
Method of Sterilization.
None IPX0 enclosure
Not splash proof. Do not spray
cleaning agents directly on
Console or Monitor enclosure.
None IPX4 enclosure (Motor)
Splash proof. Do not spray
cleaning agents directly on Motor
enclosure.
None Not for oxygen rich environment
Not suitable for use in an oxygen
rich environment.