Measurement Range | 10 to 600 mg/dL |
---|---|
Test Time | Less than 4 seconds |
Battery Life | Approximately 1000 tests |
Connectivity | Bluetooth |
Sample Size | 0.6 μL |
Weight | Approx. 40 g (1.4 oz) with battery |
Battery Type | 2 x CR2032 lithium coin cells |
Display | LCD |
Operating Conditions | 10-40°C (50-104°F), 10-90% relative humidity |
Storage Conditions | -25°C to +70°C (-13°F to +158°F) |
Describes the Accu-Chek Guide Blood Glucose Monitoring System and its intended use for single-patient diabetes management.
Outlines specific conditions and situations where the meter system should not be used for accurate results.
Provides contact information for customer support and resources for diabetes management.
Offers crucial warnings and advice regarding therapy changes based on test results and backup testing methods.
Emphasizes the biohazardous nature of the kit and warns against sharing the meter and lancing device due to infection risk.
Explains when to perform control tests and provides guidance for caregivers regarding testing and diabetes management.
Provides essential warnings and notes before beginning blood glucose testing with the system.
Details the components and features of the Accu-Chek Guide meter, including front, back, and side views.
Identifies and describes the parts of the Accu-Chek FastClix lancing device.
Explains the functions of the meter's back, arrow, and Power/Set/OK buttons.
Illustrates how to navigate through the meter's menus using display options and arrow buttons.
Lists and defines various symbols displayed on the meter screen.
Guides the user through the process of selecting the display language for the meter.
Provides instructions and notes on how to use the Accu-Chek FastClix lancing device for obtaining blood samples.
Step-by-step guide on how to load a lancet drum into the lancing device.
Details the procedure for safely removing and replacing a used lancet drum with a new one.
Provides warnings and instructions for performing blood glucose tests with the Accu-Chek Guide system.
Detailed steps for conducting a blood glucose test using blood obtained from the fingertip.
Explains how to interpret and respond to low (LO) and high (HI) blood glucose result warnings.
Guides users on how to add comments to blood glucose results for better analysis and pattern detection.
Explains the procedure and conditions for performing blood glucose tests using alternate sites like the palm or upper arm.
Provides steps to follow if blood glucose test results do not match how the user feels.
Describes the symptoms associated with hypoglycemia and hyperglycemia and actions to take.
Explains how to compare meter results with laboratory results, emphasizing fasting and timing.
Lists situations when performing a control test is recommended to ensure meter and strip accuracy.
Provides information on using and storing Accu-Chek Guide control solutions.
Step-by-step instructions for conducting a control test using the meter, test strip, and control solution.
Helps troubleshoot issues when control test results fall outside the expected range.
Introduces adjustable settings in the meter for personal preferences, including Time/Date, Beeper, Wireless, and Target Ranges.
Guides on how to set and adjust the meter's current time and date.
Explains how to turn the meter's beeper notifications on or off.
Allows users to set blood glucose target ranges for tracking results as above, within, or below range.
Explains how the meter detects and displays patterns in blood glucose results based on comments.
Details how to set up to four general test reminders per day to prompt testing.
Describes how to set reminders for testing after a meal, triggered by adding a "Before Meal" comment.
Allows selection of whether the previous blood glucose result appears with the current result.
Guides the user on how to choose the display language for the meter.
Explains how blood glucose results are stored, including capacity, deletion, and inclusion in averages.
Describes how to access and scroll through stored blood glucose results in the logbook.
Guides users on how to view average blood glucose results over different time periods.
Explains how to view the percentage of results within, above, or below target ranges over time.
Details how to track blood glucose results that fall outside target ranges, based on comments and time periods.
Explains how the meter identifies patterns in blood glucose results based on comments and target ranges.
Instructions for wirelessly transferring data from the meter to another device.
Instructions for transferring data from the meter to a PC using a USB cable and special software.
Introduces wireless synchronization of diabetes information with a mobile device and the pairing process.
Step-by-step guide for initially pairing the meter with a mobile device.
Explains how to enable or disable flight mode, which affects wireless communication availability.
Guides on selecting a default paired device for Auto-Send and Sync Time features.
Describes how to configure automatic data sending to the default paired device after each test.
Explains how to synchronize the meter's time and date with the default paired device.
Instructions for adding more paired devices to the meter, up to a limit of five.
Details the procedure for removing a previously paired device from the meter.
Identifies the approved product, Super Sani-Cloth, for cleaning and disinfecting the meter and lancing device.
Provides instructions on when, what, and how to clean and disinfect the Accu-Chek Guide meter.
Details the procedure for cleaning and disinfecting the Accu-Chek FastClix lancing device.
Step-by-step instructions for replacing the batteries in the Accu-Chek Guide meter.
Lists and explains common error messages displayed by the meter and troubleshooting steps.
Details technical specifications of the Accu-Chek Guide meter, including dimensions, power, and memory.
Explains Bluetooth technology, FCC compliance, and electromagnetic compatibility of the device.
Covers product safety standards, travel documentation, and proper disposal procedures for used materials.
Lists manual symbols and details required accessories and consumables for the system.
Outlines the 30-day money-back guarantee and the limited 3-year warranty for the Accu-Chek Guide meter.
Explains policies for service outside the warranty period and instructions for returning meters.
Details the restricted license for using the Accu-Chek Guide system, including patent information.
Outlines the restricted license for the Accu-Chek FastClix lancing device and lancet drums, including patent information.