INSTALLATION & MAINTENANCE MANUAL • X-MIND DC • (06) • 10/2018 • NXDCEN020F
Pag. 42 of 51
E44 OVERVOLTAGE ERROR CARRY OUT THE TUBEHEAD CALIBRATION
OTHERWISE, OUT OFF THE POWER SUPPLY
AND REPLACE IT
E45 ANODE VOLTAGE OUT OF TOLERANCE CARRY OUT THE TUBEHEAD CALIBRATION
OTHERWISE, OUT OFF THE POWER SUPPLY
AND REPLACE IT
E46 ANODE CURRENT OUT OF TOLERANCE CARRY OUT THE TUBEHEAD CALIBRATION
OTHERWISE, OUT OFF THE POWER SUPPLY
AND REPLACE IT
E47 CONTROL CONNECTOR CHECK THE RED AND BLACK WIRES
CONNECTION
E48 PROBLEM IN THE REFERENCE VOLTAGE OUT OFF THE POWER SUPPLY
REPLACE TIMER BOARD
ERR
MAJOR ERROR
ALL FUNCTIONS ARE DISABLED
RESET THE TIMER TO SEE THE
CODE
PLEASE NOTE
If you have any doubt for the correct solution contact immediately the manufacturer at imaging.italysupport@acteongroup.com
Please, before calling take note of the ERROR CODE.