47
ENGLISH
USER MANUAL• X-Mind trium • (13) • 03/2018 • NTR0EN010M
76 C
"Y_AXIS_ERR_RES_TOUT
Y-axis reset timeout (Uarm movement
perpendicular to the wall)"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
77 B
"R_AXIS_ERR_RES_RAMP
Unexpected FSM condition during
R-axis reset"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
78 C
"R_AXIS_ERR_RES_TOUT
R-axis reset timeout (Uarm rotation)"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
79 B
"C_AXIS_ERR_RES_RAMP
Unexpected FSM condition during
B-axis reset"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
80 C
"C_AXIS_ERR_RES_TOUT
C-axis reset timeout (CEPH image
detector slider)"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
81 B
"S_AXIS_ERR_RES_RAMP
Unexpected FSM condition during
S-axis reset"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
82 C
"S_AXIS_ERR_RES_TOUT
S-axis reset timeout (image detectors
slider on Uarm)"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
83 B
"B_AXIS_ERR_RES_RAMP
Unexpected FSM condition during
B-axis reset"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
84 C
"B_AXIS_ERR_RES_TOUT
B-axis reset timeout (bite block
vertical movement)"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
85 C
"R_AXIS_ERR_POT_BLCK
R-axis potentiometer (Uarm)
blocked (readings not coherent with
movement)"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
86 B
"R_AXIS_ERR_POT_RAMP
Unexpected FSM condition during
R-axis reset positioning based on
potentiometer value"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
87 B
"R_AXIS_ENC_HW_FATAL
Hardware error on R-axis encoder"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
88 C
"R_AXIS_ENC_CNT_ZERO
Zero-search for R-axis encoder is
failed"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
89 B
"X_AXIS_ERR_RUN_PTPT
Unexpected FSM condition during
X-axis position adjustment"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
90 B
"Y_AXIS_ERR_RUN_PTPT
Unexpected FSM condition during
Y-axis position adjustment"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
91 B
"B_AXIS_ERR_RUN_PTPT
Unexpected FSM condition during
B-axis position adjustment"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;