92 C
"R_AXIS_ERR_RUN_LIMI
R-axis position out of range"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
93 C
"HI_VOLT_ENA_IN_IDLE
XRay button activated when not
allowed"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
(Probably, someone pushed exposure button accidentally)
94 C
"INVERTER_LINK_ERROR
CAN communication error between
Mainboard and Inverter"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
95 B
"FW_UPDATED_CANT_EXE
Cannot exec rmware after update
(board need hardware reset)"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
96 C
"EEPROM_DEVICE_ERROR
Inverter board error:
GENERAL STATE ERROR, bit 0 -
EEPROM"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
97 C
"EEPROM_DMA_RD_ERROR
Inverter board error:
GENERAL STATE ERROR, bit 1 -
EEPROM READ DMA FAILURE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
98 C
"EEPROM_DMA_WR_ERROR
Inverter board error:
GENERAL STATE ERROR, bit 2 -
EEPROM WRITE DMA FAILURE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
99 C
"EEPROM_I2CBUS_ERROR
Inverter board error:
GENERAL STATE ERROR, bit 3 -
EEPROM I2CBUS"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
100 C
"ANOD_CURR_NOT_CALIB
Inverter board error:
GENERAL STATE ERROR, bit 4 -
ANODIC CURRENT NOT CALIBRATED"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
101 C
"HI_VOLT_ENA_IN_IDLE
Inverter board error:
GENERAL STATE ERROR, bit 5 - HV
ENABLE SIGNAL ACTIVE IN IDLE
MODE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
102 C
"PRE_HEAT_NOT_FINISH
Inverter board error:
GENERAL STATE ERROR, bit 6 -
FILAMENT PRE HEATING NOT
COMPLETED"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
103 C
"OVER_VOLTAGE_ON_POS
Inverter board error:
GENERAL STATE ERROR, bit 8 -
OVERVOLTAGE ON POSITIVE STAGE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
104 C
"OVER_VOLTAGE_ON_NEG
Inverter board error:
GENERAL STATE ERROR, bit 9 -
OVERVOLTAGE ON NEGATIVE STAGE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.