USER MANUAL• X-Mind trium • (13) • 03/2018 • NTR0EN010M
ENGLISH
173 B
"HORZ_ONE_RESET_RAMP
Unexpected FSM condition during
reset of primary collimator horizontal
axis"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
174 C
"HORZ_ONE_RESET_TOUT
Timeout of primary collimator
horizontal axis reset"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
175 C
"HORZ_ONE_PT_PT_GOTO
The point-to-point positioning of
horizontal axis of primary collimator
is failed"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
177 C
"XRAYS_EXAM_DO_BREAK
WARNING: operator has intentionally
stopped the Xray emission releasing
XRay button"
U
Wait the full exam procedure is completed before releasing
XRay button: message onworkstation “exposure button can
be released”