158 B
"COLUMN_POS_RUN_LIMI
Column motion stopped because of
the achievement of calibrated upper
or lower position"
U
Only a warning: operator can move in the opposite
direction.
159 C
"COLUMN_POT_REF_LIMI
Reference value for column
potentiometer is out of range"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
160 C
"COLUMN_POT_POS_LIMI
Read value of column potentiometer
is out of range"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
161 C
"COLUMN_POT_POS_BLCK
Column motion stopped because of
the column potentiometer is blocked"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
163 C
"AUX_CBCT_LINK_ERROR
CAN communication error between
Mainboard and AUX CBCT board"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
164 B
"AUX_CEPH_LINK_ERROR
CAN communication error between
Mainboard and AUX CEPH board"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
165 C
"PRM_COLLIM_DO_BREAK
WARNING: operator has intentionally
stopped a movement releasing XRay
button"
U
Wait the full exam procedure is completed before releasing
XRay button: message on workstation “exposure button can
be released”
166 C
"PRM_COLLIM_CNTL_FSM
Unexpected primary collimator FSM
condition"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
167 B
"VERT_INF_RESET_RAMP
Unexpected FSM condition during
reset of primary collimator vertical-inf
axis"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
168 C
"VERT_INF_RESET_TOUT
Timeout of primary collimator vertical-
inf axis reset"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
169 C
"VERT_INF_PT_PT_GOTO
The point-to-point positioning of
vertical-inf axis of primary collimator
is failed"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
170 B
"VERT_SUP_RESET_RAMP
Unexpected FSM condition during
reset of primary collimator vertical-
sup axis"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
171 C
"VERT_SUP_RESET_TOUT
Timeout of primary collimator vertical-
sup axis reset"
U/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
172 C
"VERT_SUP_PT_PT_GOTO
The point-to-point positioning of
vertical-sup axis of primary collimator
is failed"
A
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;