144 C
"COL_DCFI_SHORT_CIRC
Hanning board error code 13 - Short-
circuit – shut down"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
145 C
"COL_DCFI_IS_NOT_ENA
Hanning board error code 16 - Not
enabled"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
146 C
"COL_DCFI_TXCMD_TOUT
Hanning board error code 17 - Time-
out – digital interface"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
147 C
"COL_DCFI_ON_DIR_ERR
Hanning board error code 18 - Start
attempt with directional error"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
148 C
"COL_DCFI_TIMING_ERR
Hanning board error code 128 -
Internal timing error"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
149 C
"COL_DCFI_SYSTEM_ERR
Hanning board error code 129 -
System error"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
150 C
"COL_DCFI_WDOG_RESET
Hanning board error code 131 - Reset
by Watch Dog"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
151 C
"COL_DCFI_VDIP_RESET
Hanning board error code 132 - Reset
by voltage dip (brown-out)"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
152 C
"COL_DCFI_SFTW_RESET
Hanning board error code 133 - Reset
by SW"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
153 C
"COL_DCFI_E2PROM_CRC
Hanning board error code 140 -
Parameter memory CRC error"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
154 C
"COL_DCFI_E2PROM_TAB
Hanning board error code 141 -
Parameter memory table has an
error"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
155 C
"COL_DCFI_DFLT_TABLE
Hanning board error code 142 -
Parameter memory factory default
table has an error"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
156 C
"COLUMN_POS_WDG_TOUT
Column motion stopped due to CAN
communication timeout between
Mainbord and Power"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
157 C
"COLUMN_POS_END_LIMI
Column motion stopped because of
the achievement of mechanical limit
switches"
U/T
Reset error from AIS software on workstation.
Move the column in the opposite direction.
The equipment can be operated, but the occurrence of
this error means that the column potentiometer is out of
calibration; technical service is required: