USER MANUAL• X-Mind trium • (13) • 03/2018 • NTR0EN010M
129 C
"IS_OVER_LOAD_ON_NEG
Inverter board error:
GENERAL STATE ERROR, bit 38 -
OVERLOAD ON NEGATIVE HV STAGE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
130 C
"NO_FBACK_ON_POS_ERR
Inverter board error:
GENERAL STATE ERROR, bit 40 - NO
FEEDBACK ON POSITIVE HV STAGE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
131 C
"NO_FBACK_ON_NEG_ERR
Inverter board error:
GENERAL STATE ERROR, bit 41 - NO
FEEDBACK ON NEGATIVE HV STAGE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
133 C
"POWERDEV_LINK_ERROR
CAN communication error between
Mainboard and Power
"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
134 C
"COL_DCFI_LINK_ERROR
CAN communication error between
Power and Hanning board (column
board)"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
135 C
"PCF_VOLT_IS_UPP_LIM
PFC over voltage detected"
A/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
136 C
"PCF_VOLT_IS_LOW_LIM
PFC under voltage detected"
A/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
137 C
"PCF_TEMP_IS_UPP_LIM
PFC over temperature detected"
A/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
138 C
"PCF_TEMP_IS_LOW_LIM
PFC under temperature detected"
A/T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
139 D
"EMERGENCY_STOP_DONE
Trium has rebooted after releasing
the emergency stop button"
U Reset error from AIS software on workstation;
140 C
"COL_DCFI_LW_VOLTAGE
Hanning board error code 1 - Under-
voltage"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
141 C
"COL_DCFI_HI_VOLTAGE
Hanning board error code 2 - Over-
voltage"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
142 C
"COL_DCFI_HI_TEMPRAT
Hanning board error code 5 - Over-
heated motor or inverter"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
143 C
"COL_DCFI_HI_CURRENT
Hanning board error code 9 - Inverter
over-current / peak current error"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.