USER MANUAL• X-Mind trium • (13) • 03/2018 • NTR0EN010M
ENGLISH
117 C
"FILM_CURR_LOW_LIMIT
Inverter board error:
GENERAL STATE ERROR, bit 24 -
FILAMENT CURRENT TOO LOW"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
118 C
"FILM_CURR_UPP_LIMIT
Inverter board error:
GENERAL STATE ERROR, bit 25 -
FILAMENT CURRENT TOO HIGH"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
119 C
"EXP_TIME_IS_LOW_LIM
Inverter board error:
GENERAL STATE ERROR, bit 28 -
EXPOSURE TIME TOO SHORT"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
120 C
"EXP_TIME_IS_UPP_LIM
Inverter board error:
GENERAL STATE ERROR, bit 29 -
EXPOSURE TIME TOO LONG"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
121 C
"EXP_NUM_PLS_LOW_LIM
Inverter board error:
GENERAL STATE ERROR, bit 30 - TOO
FEW EXPOSURES"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
122 C
"EXP_NUM_PLS_UPP_LIM
Inverter board error:
GENERAL STATE ERROR, bit 31 - TOO
MANY EXPOSURES"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
123 C
"TUBEH_IS_NOT_DETECT
Inverter board error:
GENERAL STATE ERROR, bit 32 - RX
TUBE UNIT NOT CONNECTED OR
TEMPERATURE SENSOR FAULT"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
124 C
"TUBEH_OVER_TEMP_ERR
Inverter board error:
GENERAL STATE ERROR, bit 33 - RX
TUBE UNIT OVER TEMPERATURE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
125 C
"TUBEH_SENS_TEMP_ERR
Inverter board error:
GENERAL STATE ERROR, bit 34 - RX
TUBE UNIT TEMPERATURE SENSOR
FAULT"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
126 C
"NO_CURR_GENR_ON_POS
Inverter board error:
GENERAL STATE ERROR, bit 35 - NO
CURRENT GENERATED BY POSITIVE HV
STAGE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
127 C
"NO_CURR_GENR_ON_NEG
Inverter board error:
GENERAL STATE ERROR, bit 36 - NO
CURRENT GENERATED BY NEGATIVE
HV STAGE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.
128 C
"IS_OVER_LOAD_ON_POS
Inverter board error:
GENERAL STATE ERROR, bit 37 -
OVERLOAD ON POSITIVE HV STAGE"
T
Reset error from AIS software on workstation;
if error remains restart X-MIND trium and workstation;
If error remains technical service is required.