Instructions for use MONSOON 5.6e
20/09/2010
12/61
Maintenance Intervals, Batteries
Apparatus and/or parts must be cleaned and disinfected prior to any maintenance* - including return-
ing the apparatus for repair.
Every 6 months
- Maintenance and safety inspection according to
IEC 60601-1:1988+A1:1991+A2:1995+Cor.1995
Perform:
- Check alarm and limit functions
- Check pressure connections
- Check electrical connections
- Check safety shut-offs
Every 12 months
- Maintenance and safety inspection according to
IEC 60601-1:1988+A1:1991+A2:1995+Cor.1995
- Check alarm and limit functions
- Check pressure connections
- Check electrical connections
- Check safety shut-offs
- Calibration
- Replace the following components:
- O
2
sensor (P/N 7003)
- Qty. 2 gas supply input filter (P/N 1171)
- Water intake needle(s) (P/N 1265), qty. 2 on Double jet
- Water tube set:
- Double Jet (P/N 7918)
- Single Jet (P/N 7919)
Every 1000 operating hours
- Replace the following components:
- MATRIX Jet valve (P/N 7867), qty. 2 on Double jet
- Heating cartridge (P/N 7942)
Every 3 years replace all internal tubing
Every 10 years replace lithium battery for data backup (dispose of used battery)
Standard lithium battery CR 2032
Maintenance and safety inspection must be performed by professionals trained by ACUTRONIC
Medical Systems AG with the suitable measuring and testing devices.