ABOUT ELECTROMAGNETIC COMPATIBILITY
Note:
"MONSOON" is a MEDICAL APPLIANCE subject to specific precautionary measures with regard
to EMC and must be installed and started up according to the notices according to the instruc-
tions in this manual.
Warning:
Portable and medical HF communication devices could impact MEDICAL APPLIANCES!
Warning:
Never use "MONSOON" directly adjacent to, or stack with other apparatuses. If unavoidable, be
sure to monitor the equipment for proper operation with this set-up.
Warning:
The use other accessories, cables or converters with the "MONSOON" may result in increased
electromagnetic interference and reduce the immunity of "MONSOON".
Note:
The key performance characteristics of the "MONSOON" ventilator are:
- MONSOON must operate within the defined specifications and the medical purpose. Failure
to meet these specifications will result in respiration being stopped.
- For this reason a second, stand-alone ventilation unit must always be available, e.g. resuscita-
tion bag.
Guideline and manufacturer declaration – electromagnetic emission
The device "MONSOON" is intended for operation in the environment described below. The customer or user of the "MONSOON"
apparatus should ensure it is operated in this type of environment.
EMI measurement Compliance Electromagnetic environment - guideline
HF emission CISPR 11 Group 1
The device "MONSOON" uses HF energy solely for internal operation.
Its HF emission is thus very low and interference with adjacent appa-
ratuses is unlikely.
HF emission CISPR 11 Class A
The device "MONSOON" is suitable for use in all other areas except
residential areas and areas directly connected to a public mains supply
which also supplies buildings used for residential purposes.
Emission of harmonics IEC
61000-3-2
Class A
Emission of voltage fluctuations /
flickers IEC 61000-3-3
Met