Instructions for use MONSOON 5.6e
20/09/2010
28/61
Once the apparatus is no longer being used it
must be dehumidified to eliminate the risk of
germs potential forming inside the humidifica-
tion system.
If the power switch is activated to shut the equip-
ment off without first initiating the dehumidification
cycle, an alarm will indicate dehumidification has
not been performed.
Press the power switch again to reset the alarm
and initiate dehumidification.
The dehumidification cycle is started through the
humidification menu. The dehumidification key only
appears with the Jet switched off.
Confirm key and follow these instructions:
- Remove Jet tubes from patient
- Disconnect water supply
Be sure to keep the Jet tubes away from people.
Risk of scalding during dehumidification!
After confirming dehumidification will start and
automatically stop once the residual water has
been removed from the system. This process takes
up to 1 minute.
Dehumidification