Patient Alarms
Patient alarms are alarms triggered with (pressure) limits exceeded or disconnected water supply for
the humidification and must and can be checked by the user.
Error Cause Instructions
No drops
detected
Drop counter defective or disconnected
Bottle empty
IV set closed, kinked or disconnected
Check drop counter connection and
function
Check IV bottle
Check IV set
No water
Bottle empty
IV set closed, kinked or disconnected
Air in humidifier
Check IV bottle
Check IV set,
Press "Alarm reset"
to flood humidi-
fier
CAUTION: ventilation may stop
for up to 15 seconds
PIP high
Excessive airway pressure
Check airways
Proximal tube blocked / kinked?
Adjust ventilation parameters
PIP disconnect-
ed
No pressure change detected in proximal
tube for more than 15 seconds
Patient connected to Jet outlet?
Jet tube connected to system?
Proximal pressure sensing line
connected?
PP high
Excessive pause pressure
Check airways
Jet tube blocked / kinked?
Adjust ventilation parameter
Catheter too small for driving pres-
sure?