Instructions for use MONSOON 5.6e
25/61 20/09/2010
Superimposed Pressure Limit (SIPL)
The SIPL limit is works similar to the PIP limit but
will only block Jet valve 1 as long as the limit is
exceeded, without triggering an alarm. This meth-
od allows superimposed Jet ventilation to be im-
plemented.
The SIPL is only available on the MONSOON+ in
IPS mode (see:
Settings
).
The SIPL limit must be set to below the PP limit or
ventilation will be interrupted in response to the PP
limit.
In addition the SIPL is automatically limited by the
PIP. The following formula applies:
SIPL
max
= PIP
Is
- 5mbar
Setting the SIPL limit below 10mbar will disable this
limit; the maximum value is 40 mbar.
In graphics this SIPL limit is displayed as a broken
line!
Double Jet (Optional)
When activating the Double-Jet function the pa-
rameters for the second Jet are displayed at the
bottom right of the main screen, and can also be
adjusted from there.
In this case the parameters for bypass and humidi-
fication are no longer available. With Double-Jet
activated these settings can also be made through
the menu.