EasyManuals Logo

ACUTRONIC MONSOON User Manual

ACUTRONIC MONSOON
61 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Instructions for use MONSOON 5.6e
47/61 20/09/2010
Notices
Notices are silent and merely provide the user with information or reminders for certain processes. Notices
will only appear with a specific process active and cancel automatically.
Error Cause Instructions
Humidifier is
turned off
Safety switch-off due to mul-
tiple subsequent
No Water
Use alternate apparatus
Contact ACUTRONIC service
Check CO-
filterline if
occluded or
kinked
CO2 module pump overload-
ed
Check filter line for kink or clog, exchange filter
line if necessary
CO measuring
in progress
CO2 measurement triggered
Measurement takes approx. 20 seconds, ventila-
tion continues at prior parameters once complet-
ed
CO zeroing
Automatic calibration once
filter line is plugged into CO2
module
Do not connect filter line to patient Must be
calibrated to atmosphere
Disconnect
monitoring ac-
tive
SIPL activated
Disconnection activated
Connect proximal tube to patient
Heater and
Humidifier is
turned off
Message triggered by
Internal temperature meas-
urement defective
or
Heater error
Use alternate apparatus
Contact ACUTRONIC service
Laser ok
Appears once the oxygen
saturation preset for laser
applications is reached
Laser may be switched on

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC MONSOON and is the answer not in the manual?

ACUTRONIC MONSOON Specifications

General IconGeneral
BrandACUTRONIC
ModelMONSOON
CategoryFan
LanguageEnglish