Instructions for use MONSOON 5.6e
47/61 20/09/2010
Notices
Notices are silent and merely provide the user with information or reminders for certain processes. Notices
will only appear with a specific process active and cancel automatically.
Error Cause Instructions
Humidifier is
turned off
Safety switch-off due to mul-
tiple subsequent
No Water
Use alternate apparatus
Contact ACUTRONIC service
Check CO-
filterline if
occluded or
kinked
CO2 module pump overload-
ed
Check filter line for kink or clog, exchange filter
line if necessary
CO measuring
in progress
CO2 measurement triggered
Measurement takes approx. 20 seconds, ventila-
tion continues at prior parameters once complet-
ed
CO zeroing
Automatic calibration once
filter line is plugged into CO2
module
Do not connect filter line to patient Must be
calibrated to atmosphere
Disconnect
monitoring ac-
tive
SIPL activated
Disconnection activated
Connect proximal tube to patient
Heater and
Humidifier is
turned off
Message triggered by
Internal temperature meas-
urement defective
or
Heater error
Use alternate apparatus
Contact ACUTRONIC service
Laser ok
Appears once the oxygen
saturation preset for laser
applications is reached
Laser may be switched on