Instructions for use MONSOON 5.6e
20/09/2010
46/61
PP measuring
electronics error
Measurement discrepancy
between Main and Slave
processor
Contact ACUTRONIC service
PP sensor de-
fective
Pause pressure sensor de-
fective or disconnected
Contact ACUTRONIC service
PP sensor
pneumatically
disconnected
No pressure measurement by
sensor with Jet on
Contact ACUTRONIC service
Real Time Clock
unit error
Undefined time measurement Contact ACUTRONIC service
Low oxygen
supply pressure
O2 supply pressure too low <
2bar
Check O2 gas supply
Beeper error
Buzzer defective or loose
contact
Contact ACUTRONIC service
Ventilator error
Ventilator defective or dis-
connected
Contact ACUTRONIC service