Instructions for use MONSOON 5.6e
45/61 20/09/2010
Heater failure
Temperature too low Contact ACUTRONIC service
Internal temper-
ature measuring
defective
Temperature sensor defec-
tive or disconnected
Contact ACUTRONIC service
Jet valve error
Jet valve defective or dis-
connected
Contact ACUTRONIC service
Jet flow sensor
defective
Flow Sensor defective or
disconnected
Contact ACUTRONIC service
No communica-
tion between
MMI and main
unit
MMI unplugged
defective cable between
MMI and main unit
Check connection
Contact ACUTRONIC service
Power switch
turned OFF,
system not de-
humidified
Residual water in humidifica-
tion system
Press power switch and start dehumidification
cycle (see section:
Humidification
)
Power switch is
turned OFF. Push
‘alarm mute’ to
shutdown
Power switch on back panel
was pressed
To switch off apparatus press "Alarm mute" or
press power switch to continue operation
O2 calibration
failed
Gas supply disconnected Check gas supply and start manual calibration
O2 sensor weak
O2 sensor worn, weak signal
Replace O2 sensor or contact ACUTRONIC ser-
vice
PIP measuring
electronic error
Measurement discrepancy
between Main and Slave
processor
Contact ACUTRONIC service
PIP Sensor de-
fective
Proximal pressure sensor
defective or disconnected
Contact ACUTRONIC service