152
Corneal Reflexes exam
Tips for the patient
Patient must gaze the White led that turns on during measurement with ‘Corneal Reflexes’.
Exam procedure
The procedure is the same as refraction measurements (the single binocular and monocular
acquisitions are much faster as the Application is dedicated to eye alignment and not to
refraction measurement)
Use the top bar that gives feedback on the conditions for measurement: in particular, this
bar becomes red or green. When the bar is red, there are no conditions to acquire the
images correctly, on the contrary, when the bar is green, correct conditions for measuring
are satisfied and images can be acquired and processed correctly.
Measurement tips: usually the RED condition depends on a bad distance between the 2WIN
and the patient. When the top bar is RED adjust the distance between the 2WIN and the
patient in order to focus the image and to obtain a GREEN top bar.
The steps to perform a measurement are summarized in the following list:
1. Once the conditions for the measurement are met, hold the device in your hands and
point it towards the patient.