Chapter 8: Instrument Maintenance
OsmoPRO MAX Automated Osmometer User Guide
67
3. Select the Advanced Instruments Preventative
Maintenance checkbox to record the maintenance
event.
4. Press to enter the Advanced Instruments trained
service engineer name (required) who performed the
maintenance.
5. Press Edit to type a Comment (required - up to 1024
characters) to enter a
full explanation of the
maintenance performed
.
6. Press Save.
The instrument saves the preventative maintenance
information to the events database.
The instrument will display prompts starting 11months
after this preventative maintenance date as a
reminder that the next recommended maintenance is
due.
7. Press Exit to return to the Diagnostics menu.
Logging Repair Information (Administrators only)
Supervisor-level users can record when an instrument is
repaired and the service engineer performing the repair
by completing the following steps.
1. Press > Diagnostics > Maintenance and log in.
2. Press the Repairs tab.
3. Press to enter the Advanced Instruments trained
service engineer name (required) who repaired the
instrument.
4. Press Edit to type a Comment (required - up to 1024
characters) to explain the repair(s) performed.
5. Press Save.
6. Press Exit to return to the Diagnostics menu.
8.7 Performing a System Update
NOTE: Make sure you must have the USB drive
containing system update files.
1. Insert a USB drive into a USB port on the instrument.
2. Select > Settings > System Update and log in.
3. Select the type of update to perform: Software, Motor
Board, or Osmometer Board.
4. Press Start and wait while the system performs the
update.
A progress bar indicates the update progress.
5. When the update is finished, press Exit to return to
the Settings menu.
6. Power o the OsmoPRO MAX and wait 15 seconds.
This step is required for the system update to take
eect.
7. Power on the instrument.