EasyManuals Logo

Advanced Instruments OsmoPRO MAX User Manual

Advanced Instruments OsmoPRO MAX
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
OsmoPRO MAX Automated Osmometer User Guide
59
7.1 When to Calibrate the Instrument
NOTICE: Customers must calibrate the
OsmoPROMAX using Advanced Instruments
Calibration Standards (see Supplies, Standards,
and Controls).
Read and follow the calibration instructions
carefully. Testing calibration standards out of
order results in a calibration failure.
Calibration determines if an instrument deviates from
calibration references and, if so, to apply needed
correction factors to standardize the instrument.
Your Instrument was factory-calibrated using
2000, 850,
300, and 0
mOsm/kg H
2
O calibration standards,
Test Clinitrol™ 290 at least once daily to verify calibration.
Advanced Instruments recommends that you recalibrate
the OsmoPROMAX in any of the following cases:
The reference solution test results are out of
specification (Advanced Instruments recommends
that you test appropriate reference solutions daily to
verify the calibration)
The instrument is serviced or its hardware replaced
Calibration is required based on laboratory
regulations
NOTE: Calibration is not required after replacing the
OsmoPRO MAX Maintenance Kit.
7.2 Calibrating the Instrument
NOTICE: Read and follow these calibration
instructions carefully. Testing calibration standards
out of order results in a calibration failure.
Required Calibrators
One 2mL ampule each of 2000, 850, 300, and
0mOsm/kgH
2
O calibration standard are required to
calibrate the instrument.
These standards are included in the OsmoPRO MAX
Calibration Set
Pre-calibration Checklist
Before you begin calibrating the instrument, make sure
that the following actions are completed:
The waste bottle is empty or has sucient capacity to
hold the cleaning fluid from the test run.
There is sucient system fluid to operate the system.
Adapters to hold the ampules used for calibration
have been placed in the turntable.
Ampules of the required calibration standards and/or
controls have been opened and placed into the
adapters.
Chapter 7: Running Instrument
Calibration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoPRO MAX and is the answer not in the manual?

Advanced Instruments OsmoPRO MAX Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoPRO MAX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals