EasyManuals Logo

Advanced Instruments OsmoPRO MAX User Manual

Advanced Instruments OsmoPRO MAX
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
Appendix E: OsmoPRO MAX LIS Specification
91
OsmoPRO MAX Automated Osmometer User Guide
11.13 Requesting information status codes (optional)—
Text field containing the status of the request.
See the LIS2-A2 document for the codes and
their meaning.
Example
Q|1|A*^SP1^0|||R|20110517105358\|
DoctorXYZ|123-456-7890|Extra||N<CR>
<ETX>
Message Terminator Record
This record is used to signify the end of a message.
The fields used in this record are as specified in LIS2-A2:
12.1 Record type ID—L identifies the record as a
terminator record.
12.2 Sequence number—Sequence number (integer)
of the request.
12.3 Termination code (optional)—Single character
indicating the status of the message. See the
LIS2-A2 document for the possible codes
(defaults to N for normal if code is missing).
Example
L|1|N<CR><ETX>