Sample test procedure (with turntable installed on instrument)
To begin a batch of sample tests, press on the home screen.
Enter a user ID if it is required.
The turntable rotates and the instrument detects tubes. Filled positions appear gray.
If required, enter the Sample ID manually or scan the barcode, and press ENTER.
Load the sample tube in the load position.
Press
The turntable advances to the next tube position.
Continue loading sample tubes, or press to start testing.
NOTE: If you delay loading samples for two minutes, you are prompted to continue. You need to
respond within 30 seconds or loading cancels, and the loaded tube positions are marked empty.
Instrument Operation