Cleaning the instrument exterior : Periodically wipe the instrument with a barely-damp cloth. User either warm soapy water
or isopropyl alcohol to dampen the cloth.
Inspect sample probe daily
If sample probe shows sign of fouling, it must be cleaned by gently wiping with a lint-free cloth dampened with isopropyl alcohol or
cleaning solution for protein removal, such as Thermo Scientific Orion Cleaning Solution 900021.
Why? A dirty sample probe may affect accuracy and repeatability.
Clean cooling well weekly
Use Swab Cleaner Kit dampened with isopropyl alcohol to clean cooling well.
Why? Buildup in cooling well may affect accuracy and repeatability and may cause “Sample Did Not Freeze” errors.
Examine exhaust fans monthly
Ensure exhaust fans are unobstructed by dust or debris.
Why? Dirty exhaust fans will cause internal instrument temperature to rise, which may cause test errors or component failures.
Cleaning the Solenoid Semiannual
Instruments used daily should be cleaned monthly, while instruments that are used less frequently should be cleaned every three months.
Why? A dirty solenoid can cause “Sample Did Not Freeze” errors and can affect instrument accuracy and repeatability.
Maintenance
Refer to User Guide for more information.

Other manuals for Advanced Instruments OsmoTECH PRO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoTECH PRO and is the answer not in the manual?

Advanced Instruments OsmoTECH PRO Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoTECH PRO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals