Instrument Operation
Loading samples with the turntable removed
19
You can remove the turntable to load your samples
1. Loosen the locking screw on the center knob.
2. Lift the turntable off of the instrument and place it on a
stable location.
3. Remove the evaporation cover.
4. Load the sample tubes into positions 1 to 20.
5. Replace the wiper ring.
6. Reinstall the turntable.
7. Press .
8. Enter a user ID if it is required.
9. The turntable rotates and the instrument detects tubes.
Filled positions appear gray.
10.If the sample ID is required, enter the ID or scan the
barcode. Press ENTER. See Scanning barcodes.
11.Press .