• Interpreting Results
• Data generated using Advanced Instruments standards and reference solutions may be analyzed
according to the accuracy and precision specifications of the instrument.
Accuracy and precision (within-run repeatability) specifications apply to Advanced Instruments standards and reference solutions. Performance at Reference
Conditions: 20°C to 25°C (68°F to 77°F); 40 to 60% relative humidity.
Standards and Controls
Manufacturer’s Specifications
Accuracy • Mean Value ±3 mOsm/kg H
2
O from nominal value from 0
to 400 mOsm/kg H
2
O (1 SD)
• Mean value ±0.75% from nominal value from 400 - 2000
mOsm/kg H
2
O (1 SD)
Precision
(Within-Run
Repeatability)
• Standard deviation ≤3 mOsm/kg H
2
O from 0 to 400
mOsm/kg H
2
O
• Coefficient of variation ≤0.75% from 400 – 2000
mOsm/kg H
2
O
Refer to User Guide for more information.