After a successful 2-point calibration, you are prompted for a 3-point calibration.
1. Press YES or NO.
2. Pressing YES moves the turntable to position 7.
NOTE: For optimal performance, do not open the 2000 mOsm calibrator prior to being
prompted by the instrument.
3. Using a new, unopened vial, load three replicates of 2000 [mOsm] in
positions 7, 8, and 9.
* You can cancel the calibration at any time. The last successful calibration data
remains.
Calibration (Continued)
33