1. From the Home screen, press > CALIBRATION.
2. Log into your account.
3. Remove the turntable and any installed tubes.
4. Install a new wiper ring.
5. Load three replicates of:
• 850 [mOsm] in positions 1, 2, and 3.
• 50 [mOsm] in positions 4, 5, and 6
6. Reinstall the turntable.
7. Press NEXT.
This screen appears:
8. Press START
9. The instrument tests samples from each standard three times.
Calibration