Artifact in DR Detector images
Details An artifact is visible in the images produced by a DR
Detector.
Cause Exposure conditions have changed significantly since
latest calibration.
Brief Solution Perform calibration of the DR Detector.
For details, refer to the DX-D DR Detector Calibration
Key User Manual (0134).
DR Detector status is not changing to ready for
exposure
Details
Select the thumbnail for an exposure in the Image
Overview pane of the Examination window.
The DR Detector Switch shows the active DR Detector
and shows its status.
The DR Detector status is not changing to ready for ex-
posure.
Cause
The DR Detector is performing an internal process.
Brief Solution Wait at least 2 minutes and select the thumbnail again.
DR 10e, DR 14e, DR 17e | Problem solving | 111